Recall Z-1578-2015— Philips Ultrasound, Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Ultrasound, Inc., initiated 2015-04-10, terminated 2016-02-29. It names one FDA 510(k) clearance: K132304. Product: EPIQ 5 Ultrasound System with Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W. Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571 (1.2.2) 453561785081 (1.3) 453561800581 (1.3.1) 453561805181 (1.3.2). Reason: When Epiq 5 Ultrasound System, WITH the Pediatric Cardiology option, is set up to Metric and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patients Body Surface Area (BSA)..

Recalling firm
Philips Ultrasound, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-04-10
Terminated
2016-02-29
Firm location
Bothell, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

EPIQ 5 Ultrasound System with Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W. Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571 (1.2.2) 453561785081 (1.3) 453561800581 (1.3.1) 453561805181 (1.3.2)

Reason for recall

When Epiq 5 Ultrasound System, WITH the Pediatric Cardiology option, is set up to Metric and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patients Body Surface Area (BSA).

Data sourced from openFDA. This site is unofficial and independent of the FDA.