Recall Z-1577-2023— Diagnostica Stago, Inc.
Class II · Ongoing
Class II FDA medical device recall by Diagnostica Stago, Inc., initiated 2023-03-22. It names one FDA 510(k) clearance: K130090. Product: STA Compact Max fully automatic clinical analyzer, Reference Numbers 58602 and 58612. Reason: An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13)..
- Recalling firm
- Diagnostica Stago, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-03-22
- Firm location
- Parsippany, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
STA Compact Max fully automatic clinical analyzer, Reference Numbers 58602 and 58612
Reason for recall
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).