Recall Z-1577-2016— Medtronic

Class I · Terminated

Class I FDA medical device recall by Medtronic, initiated 2016-04-13, terminated 2017-10-17. It names one FDA 510(k) clearance: K123690. Product: Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor, Battery Packs Model Number - 016400, 010520. Battery Pack used with the Capnostream"20 and Capnostream"20p Patient Monitor. The battery pack powers the monitor when AC power is not available or when monitoring is required during intra-hospital transfer of patients. The Capnostream"20 and Capnostream"20p is a combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, and intra-hospital transport environments. Reason: Medtronic Respiratory & Monitoring Solutions Announces a Voluntary Recall of a Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor Due to the Battery Pack Might Experience Thermal Runaway with the Potential of Fire Risk to the Monitor..

Recalling firm
Medtronic
Classification
Class I
Status
Terminated
Initiated
2016-04-13
Terminated
2017-10-17
Firm location
Boulder, CO, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor, Battery Packs Model Number - 016400, 010520. Battery Pack used with the Capnostream"20 and Capnostream"20p Patient Monitor. The battery pack powers the monitor when AC power is not available or when monitoring is required during intra-hospital transfer of patients. The Capnostream"20 and Capnostream"20p is a combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, and intra-hospital transport environments

Reason for recall

Medtronic Respiratory & Monitoring Solutions Announces a Voluntary Recall of a Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor Due to the Battery Pack Might Experience Thermal Runaway with the Potential of Fire Risk to the Monitor.