Recall Z-1576-2013— Medtronic Neuromodulation
Class I · Terminated
Class I FDA medical device recall by Medtronic Neuromodulation, initiated 2013-06-03, terminated 2014-04-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic Intrathecal Catheter, model 8731SC. Contents: 38.1-cm spinal segment with length markings and guide wire, 66-cm pump segment, 15 T-gauge introducer needle, Transparent Strain-relief sleeves, Opaque Strain-relief sleeves, and V-wing anchors. Contents of inner package are STERILE. The implantable Medtronic Model 8731SC Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8731SC catheter. The catheter connects to the pump at the catheter port.. Reason: The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion..
- Recalling firm
- Medtronic Neuromodulation
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2013-06-03
- Terminated
- 2014-04-07
- Firm location
- Minneapolis, MN, United States
Product description
Medtronic Intrathecal Catheter, model 8731SC. Contents: 38.1-cm spinal segment with length markings and guide wire, 66-cm pump segment, 15 T-gauge introducer needle, Transparent Strain-relief sleeves, Opaque Strain-relief sleeves, and V-wing anchors. Contents of inner package are STERILE. The implantable Medtronic Model 8731SC Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8731SC catheter. The catheter connects to the pump at the catheter port.
Reason for recall
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
Data sourced from openFDA. This site is unofficial and independent of the FDA.