Recall Z-1574-2023— Medtronic Sofamor Danek USA Inc

Class II · Ongoing

Class II FDA medical device recall by Medtronic Sofamor Danek USA Inc, initiated 2021-11-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery. Reason: Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor..

Recalling firm
Medtronic Sofamor Danek USA Inc
Classification
Class II
Status
Ongoing
Initiated
2021-11-19
Firm location
Memphis, TN, United States

Product description

Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery

Reason for recall

Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.

Data sourced from openFDA. This site is unofficial and independent of the FDA.