Recall Z-1574-2022— Medtronic Neuromodulation

Class II · Ongoing

Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2022-07-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vanta implantable neurostimulators for pain therapy.. Reason: Potential communication issue in which the Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 may be unable to connect with a Vanta Implantable Neurostimulator (INS) Model 977006 when a patient is implanted with more than one neurostimulator..

Recalling firm
Medtronic Neuromodulation
Classification
Class II
Status
Ongoing
Initiated
2022-07-26
Firm location
Minneapolis, MN, United States

Product description

Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vanta implantable neurostimulators for pain therapy.

Reason for recall

Potential communication issue in which the Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 may be unable to connect with a Vanta Implantable Neurostimulator (INS) Model 977006 when a patient is implanted with more than one neurostimulator.