Recall Z-1573-2021— ICU Medical, Inc.
Class II · Ongoing
Class II FDA medical device recall by ICU Medical, Inc., initiated 2021-03-16. It names one FDA 510(k) clearance: K131549. Product: REF CL3946 Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock , Spinning Spiros w/Red Cap - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the administration sets needle or catheter (which is inserted into a vein).. Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position..
- Recalling firm
- ICU Medical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-03-16
- Firm location
- San Clemente, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
REF CL3946 Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock , Spinning Spiros w/Red Cap - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the administration sets needle or catheter (which is inserted into a vein).
Reason for recall
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Data sourced from openFDA. This site is unofficial and independent of the FDA.