Recall Z-1573-2013— Medtronic Neuromodulation
Class I · Terminated
Class I FDA medical device recall by Medtronic Neuromodulation, initiated 2013-06-03, terminated 2014-04-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve to be used with Medtronic SynchroMed implantable drug infusion pumps. Contents of inner package are STERILE. The Medtronic Model 8578 Sutureless Pump Connector Revision Kit is used when a pump connector for an Indura 1P Model 8709 or Model 8709SC catheter is required. The catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Indura 1P Model 8709 or Model 8709SC catheter. The catheter connects to the pump with the Model 8578 sutureless pump connector at the catheter port.. Reason: The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion..
- Recalling firm
- Medtronic Neuromodulation
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2013-06-03
- Terminated
- 2014-04-07
- Firm location
- Minneapolis, MN, United States
Product description
Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve to be used with Medtronic SynchroMed implantable drug infusion pumps. Contents of inner package are STERILE. The Medtronic Model 8578 Sutureless Pump Connector Revision Kit is used when a pump connector for an Indura 1P Model 8709 or Model 8709SC catheter is required. The catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Indura 1P Model 8709 or Model 8709SC catheter. The catheter connects to the pump with the Model 8578 sutureless pump connector at the catheter port.
Reason for recall
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
Data sourced from openFDA. This site is unofficial and independent of the FDA.