Recall Z-1572-2021— ICU Medical, Inc.

Class II · Ongoing

Class II FDA medical device recall by ICU Medical, Inc., initiated 2021-03-16. It names one FDA 510(k) clearance: K131549. Product: REF CL3538 Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock , Spinning Spiros w/Red Cap - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the administration sets needle or catheter (which is inserted into a vein).. Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2021-03-16
Firm location
San Clemente, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

REF CL3538 Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock , Spinning Spiros w/Red Cap - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the administration sets needle or catheter (which is inserted into a vein).

Reason for recall

Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.