Recall Z-1570-2026— GE Medical Systems, LLC
Class II · Ongoing
Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2026-01-30. It names one FDA 510(k) clearance: K150420. Product: Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-406, 2100169-001, 2088026-985, 2088026-971, 2088026-614, 2088026-713, 2088026-994, 2100491-006, 2088026-723, 2088026-308, 2104867-044, 2100491-012, 2088026-962, 2088026-969, 2102676-001, 2088026-936, 2088026-960, K1220JWED, 2088026-806, 2088026-506, 2088026-026, 2088026-970, 2088026-943, 2088026-990, 2088026-940, 2088026-959, 2088026-967, 2088026-989, 2104867-045, 2088026-993, 2088026-306, 2102675-010, 2088026-948, 2088026-957, 2088026-973, 2088026-946, 2088026-715, 2088026-906. Reason: There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data..
- Recalling firm
- GE Medical Systems, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-01-30
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-406, 2100169-001, 2088026-985, 2088026-971, 2088026-614, 2088026-713, 2088026-994, 2100491-006, 2088026-723, 2088026-308, 2104867-044, 2100491-012, 2088026-962, 2088026-969, 2102676-001, 2088026-936, 2088026-960, K1220JWED, 2088026-806, 2088026-506, 2088026-026, 2088026-970, 2088026-943, 2088026-990, 2088026-940, 2088026-959, 2088026-967, 2088026-989, 2104867-045, 2088026-993, 2088026-306, 2102675-010, 2088026-948, 2088026-957, 2088026-973, 2088026-946, 2088026-715, 2088026-906
Reason for recall
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Data sourced from openFDA. This site is unofficial and independent of the FDA.