Recall Z-1566-2021— ICU Medical, Inc.

Class II · Ongoing

Class II FDA medical device recall by ICU Medical, Inc., initiated 2021-03-16. It names one FDA 510(k) clearance: K081361. Product: REF CH3582 5 IN (13cm) APPX 0.57 ml, EXT SET, ANTI-SHIPHON VALVE, Spiros LOT 5125197 - Product Usage: for use in the preparation and patient administration of cytotoxic medications.. Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2021-03-16
Firm location
San Clemente, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

REF CH3582 5 IN (13cm) APPX 0.57 ml, EXT SET, ANTI-SHIPHON VALVE, Spiros LOT 5125197 - Product Usage: for use in the preparation and patient administration of cytotoxic medications.

Reason for recall

Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.

Data sourced from openFDA. This site is unofficial and independent of the FDA.