Recall Z-1566-2013— Loma Vista Medical
Class II · Terminated
Class II FDA medical device recall by Loma Vista Medical, initiated 2013-03-18, terminated 2013-07-17. It names one FDA 510(k) clearance: K121083. Product: TRUE Dilatation Balloon Valvuloplasty Catheter, REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes: 20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm; Intended to be used for balloon aortic valvuloplasty.. Reason: Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum..
- Recalling firm
- Loma Vista Medical
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-03-18
- Terminated
- 2013-07-17
- Firm location
- Burlingame, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
TRUE Dilatation Balloon Valvuloplasty Catheter, REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes: 20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm; Intended to be used for balloon aortic valvuloplasty.
Reason for recall
Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.