Recall Z-1564-2023— Bard Peripheral Vascular Inc

Class II · Ongoing

Class II FDA medical device recall by Bard Peripheral Vascular Inc, initiated 2023-04-11. It names one FDA 510(k) clearance: K171953. Product: Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM. Reason: Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure..

Recalling firm
Bard Peripheral Vascular Inc
Classification
Class II
Status
Ongoing
Initiated
2023-04-11
Firm location
Tempe, AZ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM

Reason for recall

Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.

Data sourced from openFDA. This site is unofficial and independent of the FDA.