Recall Z-1563-2023— Inpeco S.A.

Class II · Ongoing

Class II FDA medical device recall by Inpeco S.A., initiated 2023-03-10. It names one FDA 510(k) clearance: K121012. Product: Alinity h Interface Module (HSQ)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: FLX-274-20. Reason: Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences..

Recalling firm
Inpeco S.A.
Classification
Class II
Status
Ongoing
Initiated
2023-03-10
Firm location
Lugano, Switzerland

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Alinity h Interface Module (HSQ)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: FLX-274-20

Reason for recall

Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.

Data sourced from openFDA. This site is unofficial and independent of the FDA.