Recall Z-1560-2025— Baxter Healthcare Corporation

Class II · Ongoing

Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2025-03-24. It names one FDA 510(k) clearance: K141582. Product: Welch Allyn CP150 Electrocardiograph:. Reason: There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter..

Recalling firm
Baxter Healthcare Corporation
Classification
Class II
Status
Ongoing
Initiated
2025-03-24
Firm location
Deerfield, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Welch Allyn CP150 Electrocardiograph:

Reason for recall

There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

Data sourced from openFDA. This site is unofficial and independent of the FDA.