Recall Z-1560-2020— Dexcom Inc

Class II · Terminated

Class II FDA medical device recall by Dexcom Inc, initiated 2020-02-03, terminated 2022-01-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.. Reason: The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems..

Recalling firm
Dexcom Inc
Classification
Class II
Status
Terminated
Initiated
2020-02-03
Terminated
2022-01-20
Firm location
San Diego, CA, United States

Product description

Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Reason for recall

The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.