Recall Z-1559-2025— Baxter Healthcare Corporation
Class II · Ongoing
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2025-03-24. It names one FDA 510(k) clearance: K241411. Product: Welch Allyn CONNEX Accessory Power Management Stand:. Reason: There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-03-24
- Firm location
- Deerfield, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Welch Allyn CONNEX Accessory Power Management Stand:
Reason for recall
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Data sourced from openFDA. This site is unofficial and independent of the FDA.