Recall Z-1557-2014— Teleflex Medical
Class II · Terminated
Class II FDA medical device recall by Teleflex Medical, initiated 2014-03-11, terminated 2017-08-14. It names one FDA 510(k) clearance: K930738. Product: Teleflex Medical, Research Triangle Park, NC, Polypropylene, Rx Only, Sterile, Distributed by Boston Scientific, Natick, MA; CAPIO Green Braided Coated Polyglycolic Acid Suture, Synthetic Absorbable Surgical Suture SUP, RX Only, Teleflex; Deklene II Polypropylene Blue Monofil, Blue Monofilament Polypropylene Suture, Nonabsorbable Surgical Suture USP, Deknatel; Deklene MAXX, Blue Monofilament Polypropylene suture Nonabsorbable Surgical Suture USP, Deknatel. Indicated for use in soft tissue approximation.. Reason: Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use..
- Recalling firm
- Teleflex Medical
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-03-11
- Terminated
- 2017-08-14
- Firm location
- Durham, NC, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Teleflex Medical, Research Triangle Park, NC, Polypropylene, Rx Only, Sterile, Distributed by Boston Scientific, Natick, MA; CAPIO Green Braided Coated Polyglycolic Acid Suture, Synthetic Absorbable Surgical Suture SUP, RX Only, Teleflex; Deklene II Polypropylene Blue Monofil, Blue Monofilament Polypropylene Suture, Nonabsorbable Surgical Suture USP, Deknatel; Deklene MAXX, Blue Monofilament Polypropylene suture Nonabsorbable Surgical Suture USP, Deknatel. Indicated for use in soft tissue approximation.
Reason for recall
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.