Recall Z-1552-2026— Beckman Coulter Mishima K.K.

Class II · Ongoing

Class II FDA medical device recall by Beckman Coulter Mishima K.K., initiated 2026-01-30. FDA's recall record names no specific 510(k) clearance for this event. Product: DxC 700 AU, REF: B86444, B86446. Reason: A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment..

Recalling firm
Beckman Coulter Mishima K.K.
Classification
Class II
Status
Ongoing
Initiated
2026-01-30
Firm location
Sunto-Gun, N/A, Japan

Product description

DxC 700 AU, REF: B86444, B86446

Reason for recall

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.