Recall Z-1551-2015— Fresenius Medical Care Holdings, Inc.
Class II · Terminated
Class II FDA medical device recall by Fresenius Medical Care Holdings, Inc., initiated 2015-02-20, terminated 2021-05-18. It names one FDA 510(k) clearance: K994267. Product: 2008 Series Hemodialysis Machines: 2008K with the following product code and description: 190305 2008K (w/ OLC); 190371 2008K System Std (w/OLC and Diasafe Plus); 190300 2008K System Std (w/OLC and Diasafe); 190336 2008K System w/ Fifth Module Holder; 190517 2008K System w/ OLC and Diasafe Plus - Spanish; 190302 2008K System w/ OLC and Short Cab; 190306 2008K System w/ Short Cabinet; 190373 2008K System w/ Short Cabinet; 190305 2008K System w/o Diasafe; 190301 2008K System w/o HP; 190372 2008K System w/o HP; 190303 2008K System w/o OLC and Diasafe; 190304 2008K System w/o OLC and Diasafe and HP; 190452 2008K System w/o OLC and w/ Diasafe Plus; 190303 2008K W/O OLC and DIASAFE W/ HEPARIN PUMP; 190304 2008K W/O OLC and DIASAFE W/O HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.. Reason: The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen..
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-02-20
- Terminated
- 2021-05-18
- Firm location
- Waltham, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
2008 Series Hemodialysis Machines: 2008K with the following product code and description: 190305 2008K (w/ OLC); 190371 2008K System Std (w/OLC and Diasafe Plus); 190300 2008K System Std (w/OLC and Diasafe); 190336 2008K System w/ Fifth Module Holder; 190517 2008K System w/ OLC and Diasafe Plus - Spanish; 190302 2008K System w/ OLC and Short Cab; 190306 2008K System w/ Short Cabinet; 190373 2008K System w/ Short Cabinet; 190305 2008K System w/o Diasafe; 190301 2008K System w/o HP; 190372 2008K System w/o HP; 190303 2008K System w/o OLC and Diasafe; 190304 2008K System w/o OLC and Diasafe and HP; 190452 2008K System w/o OLC and w/ Diasafe Plus; 190303 2008K W/O OLC and DIASAFE W/ HEPARIN PUMP; 190304 2008K W/O OLC and DIASAFE W/O HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Reason for recall
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Data sourced from openFDA. This site is unofficial and independent of the FDA.