Recall Z-1550-2025— CareFusion 303, Inc.
Class II · Ongoing
Class II FDA medical device recall by CareFusion 303, Inc., initiated 2025-03-11. FDA's recall record names no specific 510(k) clearance for this event. Product: BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 155143-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 155144-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 155145-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 155157-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P / 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 155153-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 155155-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 155147-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 155149-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 155150-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 169-114 BD PYXIS MEDBANK MINI CR-4HH-P / 169-115 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 169-83 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 169-85 BD PYXIS MEDBANK TWR MN CR-14HH-P / 169-86 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 169-89 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 169-91 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /169-93 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 169-95 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 169-99 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 138913-01 BD PYXIS MEDBANK MINI CR-4HH-P / 138911-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P / 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P / 139002-01 BD PYXIS MEDBANK MN 200 CR-4HH-P / 139000-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P / 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P / 139018-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P / 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P / 139021-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P / 139022-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 138975-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 138976-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 138977-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 139085-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 138964-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 138965-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 138970-01 BD Pyxis QFill Replenishment Station / 138904-01 KIT BD PYXIS QFILL REPLENISHMENT STATION / 155197-01 MEDBANK UPOS211 AIO WIN10 QFILL 1.5 / 155215-05 OPT,MEDBANK,HD,WIN10,QFILL,CR / 131944-04 The BD Pyxis" MedBank" CUBIE" Replenishment Station is an Automated Dispensing Cabinet (ADC) that is intended to be used by a qualified and authorized user to replenish medications and supplies to a BD Pyxis" MedBank" cabinet.. Reason: Due to a software issue, when attempting to restock the automated dispensing cabinet, user have received the banner message "Stocking window has passed, or item has been removed from purchase order. Please contact Pharmacy for assistance". When this occurs, the medication cannot be restocked and can potentially lead to delays in accessing medication for patients..
- Recalling firm
- CareFusion 303, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-03-11
- Firm location
- San Diego, CA, United States
Product description
BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 155143-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 155144-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 155145-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 155157-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P / 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 155153-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 155155-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 155147-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 155149-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 155150-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 169-114 BD PYXIS MEDBANK MINI CR-4HH-P / 169-115 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 169-83 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 169-85 BD PYXIS MEDBANK TWR MN CR-14HH-P / 169-86 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 169-89 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 169-91 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /169-93 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 169-95 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 169-99 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 138913-01 BD PYXIS MEDBANK MINI CR-4HH-P / 138911-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P / 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P / 139002-01 BD PYXIS MEDBANK MN 200 CR-4HH-P / 139000-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P / 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P / 139018-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P / 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P / 139021-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P / 139022-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 138975-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 138976-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 138977-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 139085-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 138964-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 138965-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 138970-01 BD Pyxis QFill Replenishment Station / 138904-01 KIT BD PYXIS QFILL REPLENISHMENT STATION / 155197-01 MEDBANK UPOS211 AIO WIN10 QFILL 1.5 / 155215-05 OPT,MEDBANK,HD,WIN10,QFILL,CR / 131944-04 The BD Pyxis" MedBank" CUBIE" Replenishment Station is an Automated Dispensing Cabinet (ADC) that is intended to be used by a qualified and authorized user to replenish medications and supplies to a BD Pyxis" MedBank" cabinet.
Reason for recall
Due to a software issue, when attempting to restock the automated dispensing cabinet, user have received the banner message "Stocking window has passed, or item has been removed from purchase order. Please contact Pharmacy for assistance". When this occurs, the medication cannot be restocked and can potentially lead to delays in accessing medication for patients.
Data sourced from openFDA. This site is unofficial and independent of the FDA.