Recall Z-1547-2013— Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Biomet, Inc., initiated 2013-04-29, terminated 2015-05-06. FDA's recall record names no specific 510(k) clearance for this event. Product: Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.. Reason: Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su.

Recalling firm
Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-04-29
Terminated
2015-05-06
Firm location
Warsaw, IN, United States

Product description

Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.

Reason for recall

Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su

Data sourced from openFDA. This site is unofficial and independent of the FDA.