Recall Z-1546-2026— Trividia Health, Inc.
Class I · Ongoing
Class I FDA medical device recall by Trividia Health, Inc., initiated 2026-02-06. It names one FDA 510(k) clearance: K140100. Product: Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.. Reason: The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose..
- Recalling firm
- Trividia Health, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2026-02-06
- Firm location
- Fort Lauderdale, FL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.
Reason for recall
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.