Recall Z-1541-2024— Smiths Medical ASD Inc.

Class II · Ongoing

Class II FDA medical device recall by Smiths Medical ASD Inc., initiated 2023-12-19. It names one FDA 510(k) clearance: K111386. Product: Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4000-0105-78, h) 4000-0106-00, i) 4000-0106-01, j) 4000-0106-231,. Reason: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded..

Recalling firm
Smiths Medical ASD Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-12-19
Firm location
Minneapolis, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4000-0105-78, h) 4000-0106-00, i) 4000-0106-01, j) 4000-0106-231,

Reason for recall

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

Data sourced from openFDA. This site is unofficial and independent of the FDA.