Recall Z-1541-2020— Siemens Healthcare Diagnostics, Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2020-02-13, terminated 2021-04-23. It names one FDA 510(k) clearance: K061719. Product: Dimension Vista¿ Total Bilirubin Flex¿ reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10445146 Product Usage: Dimension Vista TBIL method is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma on the Dimension Vista¿ System. Measurements of total bilirubin are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder disease.. Reason: Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy.
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-02-13
- Terminated
- 2021-04-23
- Firm location
- Newark, DE, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Dimension Vista¿ Total Bilirubin Flex¿ reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10445146 Product Usage: Dimension Vista TBIL method is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma on the Dimension Vista¿ System. Measurements of total bilirubin are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder disease.
Reason for recall
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy