Recall Z-1540-2020— Siemens Healthcare Diagnostics, Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2020-02-13, terminated 2021-04-23. It names one FDA 510(k) clearance: K060628. Product: Dimension Total Bilirubin Flex reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10444957 Product Usage: Dimension TBI method for the ¿ clinical chemistry syst6em is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma. Dimension Measurements of total bilirubin are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder disease.. Reason: Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy.
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-02-13
- Terminated
- 2021-04-23
- Firm location
- Newark, DE, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Dimension Total Bilirubin Flex reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10444957 Product Usage: Dimension TBI method for the ¿ clinical chemistry syst6em is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma. Dimension Measurements of total bilirubin are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder disease.
Reason for recall
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy