Recall Z-1538-2012— Philips Healthcare Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Healthcare Inc., initiated 2011-10-19, terminated 2017-01-18. FDA's recall record names no specific 510(k) clearance for this event. Product: Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide. Reason: When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie.

Recalling firm
Philips Healthcare Inc.
Classification
Class II
Status
Terminated
Initiated
2011-10-19
Terminated
2017-01-18
Firm location
Andover, MA, United States

Product description

Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide

Reason for recall

When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie