Recall Z-1536-2021— Fresenius Medical Care Holdings, Inc.
Class II · Terminated
Class II FDA medical device recall by Fresenius Medical Care Holdings, Inc., initiated 2021-03-18, terminated 2024-04-10. It names one FDA 510(k) clearance: K133829. Product: AquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. Fresenius P/N 24-2500-0 All 3 units make up the AquaBplus 2500Fresenius Part Number: G02040109-US AquaBplus 3000 - Typ USA G02040115-US AquaBplus 2500 - Typ USA G02040120FMC-US AquaBplus HF- Typ USA. Reason: AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation.
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-03-18
- Terminated
- 2024-04-10
- Firm location
- Waltham, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
AquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. Fresenius P/N 24-2500-0 All 3 units make up the AquaBplus 2500Fresenius Part Number: G02040109-US AquaBplus 3000 - Typ USA G02040115-US AquaBplus 2500 - Typ USA G02040120FMC-US AquaBplus HF- Typ USA
Reason for recall
AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Data sourced from openFDA. This site is unofficial and independent of the FDA.