Recall Z-1533-2021— GE Healthcare, LLC
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2021-03-19, terminated 2023-08-11. It names one FDA 510(k) clearance: K182419. Product: Centricity Universal Viewer - Product Usage: intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.. Reason: When the Cross Enterprise Display option is enabled on the Centricity Universal Viewer and Workflow Manager option is used as the PACS worklist launching the viewer, the patient jacket may potentially include a different patient s comparison exam(s)..
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-03-19
- Terminated
- 2023-08-11
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Centricity Universal Viewer - Product Usage: intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.
Reason for recall
When the Cross Enterprise Display option is enabled on the Centricity Universal Viewer and Workflow Manager option is used as the PACS worklist launching the viewer, the patient jacket may potentially include a different patient s comparison exam(s).