Recall Z-1529-2016— Mevion Medical Systems, Inc.
Class II · Terminated
Class II FDA medical device recall by Mevion Medical Systems, Inc., initiated 2016-03-15, terminated 2017-08-24. It names one FDA 510(k) clearance: K120676. Product: MEVION S250 for Proton Radiation Therapy. Reason: Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death..
- Recalling firm
- Mevion Medical Systems, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-03-15
- Terminated
- 2017-08-24
- Firm location
- Littleton, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MEVION S250 for Proton Radiation Therapy
Reason for recall
Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death.
Data sourced from openFDA. This site is unofficial and independent of the FDA.