Recall Z-1528-2021— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Class II · Ongoing

Class II FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2021-04-06. FDA's recall record names no specific 510(k) clearance for this event. Product: CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A. Reason: A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted..

Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Classification
Class II
Status
Ongoing
Initiated
2021-04-06
Firm location
Mounds View, MN, United States

Product description

CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A

Reason for recall

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.