Recall Z-1527-2018— Covidien LLC
Class II · Terminated
Class II FDA medical device recall by Covidien LLC, initiated 2018-04-09, terminated 2020-05-12. It names one FDA 510(k) clearance: K083781. Product: EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology 3.5MM, Product Number HEM3335, GTIN Number 1 - 10884521047518, GTIN Number 2 - 20884521047515 Product Usage: The EEA Hemorrhoid and prolapse stapler set with DST Series Technology (instrument) is designed for use as a stapler instrument for control of rectal prolapse and hemorrhoid disease. The instrument places a circular, double-staggered row of titanium DST staples and removes a circular tissue specimen. The instrument is offered in a 33mm diameter size with either 3.5mm or 4.8mm staples. The accompanying accessories comprised of an access port, anoscope, and dilator. The access port aids the insertion and removal of the anoscope and the stapler device by dilating the anal sphincter throughout the procedure. The anoscope is designed to facilitate pursestring suture application above the dentate line. It is designed with a guide and markings to assist in creating a circumferential pursestring suture at a consistent placement within the rectum. The dilator is designed to aid the initial insertion of the anoscope and port by retracting the rectal tissue. Both the anoscope and port are transparent to facilitate visualization of the anal canal. The EEA Hemorrhoid and prolapse stapler set with DST Series Technology has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease. The EEA Hemorrhoid and prolapse stapler also has application in the distal alimentary tract for the creation of end-to-end and end-to-side anastomoses.. Reason: There is a potential for improper welding of the yellow staple guide to the instrument. Use of a device with an improperly welded staple guide may result in improper staple formation potentially leading to bleeding or anastomotic leak..
- Recalling firm
- Covidien LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-04-09
- Terminated
- 2020-05-12
- Firm location
- North Haven, CT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology 3.5MM, Product Number HEM3335, GTIN Number 1 - 10884521047518, GTIN Number 2 - 20884521047515 Product Usage: The EEA Hemorrhoid and prolapse stapler set with DST Series Technology (instrument) is designed for use as a stapler instrument for control of rectal prolapse and hemorrhoid disease. The instrument places a circular, double-staggered row of titanium DST staples and removes a circular tissue specimen. The instrument is offered in a 33mm diameter size with either 3.5mm or 4.8mm staples. The accompanying accessories comprised of an access port, anoscope, and dilator. The access port aids the insertion and removal of the anoscope and the stapler device by dilating the anal sphincter throughout the procedure. The anoscope is designed to facilitate pursestring suture application above the dentate line. It is designed with a guide and markings to assist in creating a circumferential pursestring suture at a consistent placement within the rectum. The dilator is designed to aid the initial insertion of the anoscope and port by retracting the rectal tissue. Both the anoscope and port are transparent to facilitate visualization of the anal canal. The EEA Hemorrhoid and prolapse stapler set with DST Series Technology has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease. The EEA Hemorrhoid and prolapse stapler also has application in the distal alimentary tract for the creation of end-to-end and end-to-side anastomoses.
Reason for recall
There is a potential for improper welding of the yellow staple guide to the instrument. Use of a device with an improperly welded staple guide may result in improper staple formation potentially leading to bleeding or anastomotic leak.
Data sourced from openFDA. This site is unofficial and independent of the FDA.