Recall Z-1523-2015— Zimmer Gmbh

Class II · Terminated

Class II FDA medical device recall by Zimmer Gmbh, initiated 2015-03-18, terminated 2015-09-23. FDA's recall record names no specific 510(k) clearance for this event. Product: Anatomical Shoulder Handle for Rasp Item: 01.04233.000 Used during total joint arthroplasty of the shoulder.. Reason: Complaints of difficulties to attach the Anatomical Shoulder Rasps (high resistance) or, once attached, due to difficulties to remove the handle (seizing up of the two components). This has the potential for delay in surgery..

Recalling firm
Zimmer Gmbh
Classification
Class II
Status
Terminated
Initiated
2015-03-18
Terminated
2015-09-23
Firm location
Winterthur, Switzerland

Product description

Anatomical Shoulder Handle for Rasp Item: 01.04233.000 Used during total joint arthroplasty of the shoulder.

Reason for recall

Complaints of difficulties to attach the Anatomical Shoulder Rasps (high resistance) or, once attached, due to difficulties to remove the handle (seizing up of the two components). This has the potential for delay in surgery.