Recall Z-1522-2021— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Class II · Ongoing

Class II FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2021-04-06. FDA's recall record names no specific 510(k) clearance for this event. Product: CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).. Reason: There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT)..

Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Classification
Class II
Status
Ongoing
Initiated
2021-04-06
Firm location
Mounds View, MN, United States

Product description

CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).

Reason for recall

There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).