Recall Z-1520-2013— DePuy Orthopaedics, Inc.

Class II · Terminated

Class II FDA medical device recall by DePuy Orthopaedics, Inc., initiated 2013-05-03, terminated 2013-08-28. FDA's recall record names no specific 510(k) clearance for this event. Product: DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.. Reason: The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant..

Recalling firm
DePuy Orthopaedics, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-05-03
Terminated
2013-08-28
Firm location
Warsaw, IN, United States

Product description

DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.

Reason for recall

The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant.