Recall Z-1519-2021— Oto Med Inc
Class II · Ongoing
Class II FDA medical device recall by Oto Med Inc, initiated 2020-12-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Grace Medical REF S-2500 *** SHEEHY BONE DUST COLLECTOR *** Stainless steel filter - Product Usage: intended to be inserted into a suction tube during the early stages of otologic surgery to collect bone particles for future use.. Reason: During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices..
- Recalling firm
- Oto Med Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-12-07
- Firm location
- Lake Havasu City, AZ, United States
Product description
Grace Medical REF S-2500 *** SHEEHY BONE DUST COLLECTOR *** Stainless steel filter - Product Usage: intended to be inserted into a suction tube during the early stages of otologic surgery to collect bone particles for future use.
Reason for recall
During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.