Recall Z-1518-2024— Intuitive Surgical, Inc.

Class II · Ongoing

Class II FDA medical device recall by Intuitive Surgical, Inc., initiated 2024-03-13. It names 2 FDA 510(k) clearances: K182371, K173906. Product: Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098. Reason: Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient..

Recalling firm
Intuitive Surgical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-03-13
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

Reason for recall

Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.

Data sourced from openFDA. This site is unofficial and independent of the FDA.