Recall Z-1517-2025— Beckman Coulter Inc.

Class II · Ongoing

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2025-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.. Reason: Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the operator at the analyzer or automatically triggered by predefined rules in System Configuration, the calculated result will not be recalculated using the constituent tests rerun results. The system will report only the calculated result using the initial constituent tests results to the operator and LIS/Remisol. The issue was identified by Beckman Coulter internal engineering testing. The defect may cause a delay in reporting patient results..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-02-26
Firm location
Brea, CA, United States

Product description

DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Reason for recall

Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the operator at the analyzer or automatically triggered by predefined rules in System Configuration, the calculated result will not be recalculated using the constituent tests rerun results. The system will report only the calculated result using the initial constituent tests results to the operator and LIS/Remisol. The issue was identified by Beckman Coulter internal engineering testing. The defect may cause a delay in reporting patient results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.