Recall Z-1514-2020— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2020-02-05, terminated 2021-11-18. FDA's recall record names no specific 510(k) clearance for this event. Product: IMMULITE 2000 Systems E2 Estradiol Catalog #: US: L2KE22(D) (200 tests) and L2KE26(D) (600 tests) and OUS: L2KE22 (200 tests), L2KE26 (600 tests) - Product Usage: For in vitro diagnostic use with the IMMULITE systems Analyzers for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders.. Reason: Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-02-05
Terminated
2021-11-18
Firm location
Tarrytown, NY, United States

Product description

IMMULITE 2000 Systems E2 Estradiol Catalog #: US: L2KE22(D) (200 tests) and L2KE26(D) (600 tests) and OUS: L2KE22 (200 tests), L2KE26 (600 tests) - Product Usage: For in vitro diagnostic use with the IMMULITE systems Analyzers for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders.

Reason for recall

Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer.

Data sourced from openFDA. This site is unofficial and independent of the FDA.