Recall Z-1513-2014— Brainlab AG

Class II · Terminated

Class II FDA medical device recall by Brainlab AG, initiated 2014-02-24, terminated 2017-04-24. It names one FDA 510(k) clearance: K120789. Product: ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED. Reason: When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position..

Recalling firm
Brainlab AG
Classification
Class II
Status
Terminated
Initiated
2014-02-24
Terminated
2017-04-24
Firm location
Feldkirchen, Germany

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED

Reason for recall

When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.

Data sourced from openFDA. This site is unofficial and independent of the FDA.