Recall Z-1511-2022— GE Healthcare, LLC

Class II · Ongoing

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2022-06-02. FDA's recall record names no specific 510(k) clearance for this event. Product: Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems.. Reason: As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Ongoing
Initiated
2022-06-02
Firm location
Waukesha, WI, United States

Product description

Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems.

Reason for recall

As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.