Recall Z-1510-2017— Ohmeda Medical

Class II · Terminated

Class II FDA medical device recall by Ohmeda Medical, initiated 2016-11-18, terminated 2018-07-13. It names one FDA 510(k) clearance: K974349. Product: CarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most incubators can be used in two operating modes: 1. Air Control: The clinician sets the appropriate air temperature for maintaining the desired patient temperature. The air temperature is initially selected based on the clinician s training and experience and then is adjusted based on the patient s needs and clinical status. 2. Patient control: The clinician sets the desired patient temperature. A skin temperature probe sense the patient temperature and feeds this information to the controller of the incubator. Thecontroller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Incubators have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Incubators may incorporate other features, such as humidification of the infant environment, tilting of the bed, oxygen supply, and data output to remote monitors or nurse call systems. Reason: Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury..

Recalling firm
Ohmeda Medical
Classification
Class II
Status
Terminated
Initiated
2016-11-18
Terminated
2018-07-13
Firm location
Laurel, MD, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most incubators can be used in two operating modes: 1. Air Control: The clinician sets the appropriate air temperature for maintaining the desired patient temperature. The air temperature is initially selected based on the clinician s training and experience and then is adjusted based on the patient s needs and clinical status. 2. Patient control: The clinician sets the desired patient temperature. A skin temperature probe sense the patient temperature and feeds this information to the controller of the incubator. Thecontroller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Incubators have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Incubators may incorporate other features, such as humidification of the infant environment, tilting of the bed, oxygen supply, and data output to remote monitors or nurse call systems

Reason for recall

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.