Recall Z-1508-2020— Stryker Corporation
Class II · Terminated
Class II FDA medical device recall by Stryker Corporation, initiated 2019-09-12, terminated 2022-10-31. It names one FDA 510(k) clearance: K963765. Product: Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electrocautery Probe, J Tip, 5MM 0250070443-PKG, StrykeFlow Electrocautery Probe, L Tip, 5MM 0250070444-PKG, StrykeFlow Electrocautery Probe, Ball Tip, 5MM 0250070445-PKG, StrykeFlow Electrocautery Probe, Needle Tip 5MM 0250070446-PKG, StrykeFlow Electrocautery Probe, Spoon Tip, 5MM. Reason: Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately..
- Recalling firm
- Stryker Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-09-12
- Terminated
- 2022-10-31
- Firm location
- San Jose, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electrocautery Probe, J Tip, 5MM 0250070443-PKG, StrykeFlow Electrocautery Probe, L Tip, 5MM 0250070444-PKG, StrykeFlow Electrocautery Probe, Ball Tip, 5MM 0250070445-PKG, StrykeFlow Electrocautery Probe, Needle Tip 5MM 0250070446-PKG, StrykeFlow Electrocautery Probe, Spoon Tip, 5MM
Reason for recall
Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.