Recall Z-1507-2024— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2020-12-24. It names one FDA 510(k) clearance: K180015. Product: PHILIPS Incisive Computed tomography X-ray system Incisive CT Software Upgrade from V4.5.1.41921 to V4.5.1.41989-Intended to produce cross-sectional images of the body Model Number: 728143. Reason: Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured.

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2020-12-24
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

PHILIPS Incisive Computed tomography X-ray system Incisive CT Software Upgrade from V4.5.1.41921 to V4.5.1.41989-Intended to produce cross-sectional images of the body Model Number: 728143

Reason for recall

Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured