Recall Z-1506-2022— Guangdong Haiou Medical Apparatus Co., Ltd

Class II · Ongoing

Class II FDA medical device recall by Guangdong Haiou Medical Apparatus Co., Ltd, initiated 2022-06-10. FDA's recall record names no specific 510(k) clearance for this event. Product: HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25. Reason: Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user.

Recalling firm
Guangdong Haiou Medical Apparatus Co., Ltd
Classification
Class II
Status
Ongoing
Initiated
2022-06-10
Firm location
Puning, China

Product description

HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25

Reason for recall

Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user

Data sourced from openFDA. This site is unofficial and independent of the FDA.