Class II FDA medical device recall by Alcon Grieshaber AG, initiated 2013-04-17, terminated 2015-01-07. FDA's recall record names no specific 510(k) clearance for this event. Product: The Constellation Vision System Pneumatic Handpiece. Indicated for both anterior and posterior segment opthalmic surgery.. Reason: Alcon is conducting a Medical Device Removal for the Constellation Pneumatic Handpiece, DSP, product number 725.01. This action is being taken because the forceps or scissors tips (sold separately) may dislodge from the hand piece, if not attached by the user according to the Directions for Use..

Recalling firm
Alcon Grieshaber AG
Classification
Class II
Status
Terminated
Initiated
2013-04-17
Terminated
2015-01-07
Firm location
Schaffhausen, N/A, Switzerland

Product description

The Constellation Vision System Pneumatic Handpiece. Indicated for both anterior and posterior segment opthalmic surgery.

Reason for recall

Alcon is conducting a Medical Device Removal for the Constellation Pneumatic Handpiece, DSP, product number 725.01. This action is being taken because the forceps or scissors tips (sold separately) may dislodge from the hand piece, if not attached by the user according to the Directions for Use.