Recall Z-1503-2014— Philips Medical Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems, Inc., initiated 2014-03-17, terminated 2020-09-16. It names one FDA 510(k) clearance: K982795. Product: Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85. Reason: The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held..

Recalling firm
Philips Medical Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-03-17
Terminated
2020-09-16
Firm location
Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85

Reason for recall

The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.