Recall Z-1500-2016— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2016-03-10, terminated 2018-06-07. It names one FDA 510(k) clearance: K111534. Product: ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD The ADVIA 560 Hematology System is a fully-automated, high-quality hematology system for in vitro diagnostic use in clinical laboratories. Reason: Software anomaly; Siemens identified that software version 1.4.2133 on the ADVIA 560 Hematology System does not trigger the following flags: The G or L morphology flags for immature granulocytes (IG) and atypical lymphocytes (ATYP), respectively. The results from patient samples which have immature granulocytes or atypical lymphocytes will not generate the flags when they should. As defined in t.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-03-10
Terminated
2018-06-07
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD The ADVIA 560 Hematology System is a fully-automated, high-quality hematology system for in vitro diagnostic use in clinical laboratories

Reason for recall

Software anomaly; Siemens identified that software version 1.4.2133 on the ADVIA 560 Hematology System does not trigger the following flags: The G or L morphology flags for immature granulocytes (IG) and atypical lymphocytes (ATYP), respectively. The results from patient samples which have immature granulocytes or atypical lymphocytes will not generate the flags when they should. As defined in t