Recall Z-1499-2016— Leica Microsystems, Inc.

Class II · Terminated

Class II FDA medical device recall by Leica Microsystems, Inc., initiated 2016-01-11, terminated 2017-04-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Leica Microsystems (Schweiz) Ag. The common name of the device is surgical microscope. It has been assigned the product code FSO by the FDA and is classified as a Class I medical device. Intended for improving the visibility of objects through magnification and illumination.. Reason: An electronic component of the Leica M525 OH4 and Leica M720 OH5 surgical microscopes with 400W illumination could potentially overheat, resulting in smoke and smell emanating from the ventilation openings of the illumination units located opposite and away from the surgeon / patient side of the surgical microscope..

Recalling firm
Leica Microsystems, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-01-11
Terminated
2017-04-24
Firm location
Buffalo Grove, IL, United States

Product description

Leica Microsystems (Schweiz) Ag. The common name of the device is surgical microscope. It has been assigned the product code FSO by the FDA and is classified as a Class I medical device. Intended for improving the visibility of objects through magnification and illumination.

Reason for recall

An electronic component of the Leica M525 OH4 and Leica M720 OH5 surgical microscopes with 400W illumination could potentially overheat, resulting in smoke and smell emanating from the ventilation openings of the illumination units located opposite and away from the surgeon / patient side of the surgical microscope.

Data sourced from openFDA. This site is unofficial and independent of the FDA.