Recall Z-1498-2023— Siemens Medical Solutions USA, Inc

Class II · Ongoing

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2023-02-02. It names one FDA 510(k) clearance: K163286. Product: ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000. Reason: If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the C-Arm will reach the target position with an inaccuracy of 5-10 mm. As a result, the message Endposition reached will be displayed .Leading to unintended direction of the movement causing crushing of a patient, staff member, operator, or equipmen Live images may not match the previously stored reference images. Overlay images may be shown inaccurate on anatomy (e.g., DSA Roadmap workflow does not match real anatomy). This may cause e.g., a vessel perforation in DSA roadmap..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Ongoing
Initiated
2023-02-02
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000

Reason for recall

If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the C-Arm will reach the target position with an inaccuracy of 5-10 mm. As a result, the message Endposition reached will be displayed .Leading to unintended direction of the movement causing crushing of a patient, staff member, operator, or equipmen Live images may not match the previously stored reference images. Overlay images may be shown inaccurate on anatomy (e.g., DSA Roadmap workflow does not match real anatomy). This may cause e.g., a vessel perforation in DSA roadmap.